CLIA 2567s by specialty: how deficiencies are written up, and how labs answer them
CLIA 2567s follow the same structure in every specialty: a D-tag, the 42 CFR 493 citation, the requirement, a statement that it was not met, and the evidence. What changes by specialty is which requirements get cited. This library groups common finding patterns by laboratory area, with the regulation behind each and what a complete correction covers.

Examples on this site are illustrative composites written to show how findings are typically structured. They are not quotations from any specific laboratory's CMS-2567.
Browse by laboratory area
Chemistry and hematology
Test system conditions, calibration, QC review, reagent dating and performance verification.
Microbiology and molecular
Media and reagent QC, procedure manuals, contamination control and method verification.
Waived and PPM testing
Manufacturer instructions, certificate scope and provider-performed microscopy.
Personnel and proficiency testing
Director duties, competency, qualifications, PT enrollment and PT handling.
Find real public 2567s
How to search CMS's QCOR system for published CLIA statements of deficiencies.
The most-cited tags across all specialties
CMS's Year 2024 Top 10 list shows which standard-level tags were cited at the most laboratories: D5413 (test system conditions, §493.1252(b)) at 818 labs, D5209 (competency assessment policies, §493.1235) at 778, D5401 (procedure manual followed, §493.1251(a)) at 635, D5403 (procedure manual content) at 609, D5417 (expired or deteriorated materials) at 607, D5217 (twice-yearly accuracy checks, §493.1236(c)(1)) at 548, D5421 (performance verification, §493.1253(b)(1)) at 451, D6016 (director PT responsibilities) at 451, D5429 (maintenance) at 449 and D5805 (test report content, §493.1291(c)) at 417.
Why specialty context matters
The same tag reads differently in different departments. D5413 in a chemistry lab is usually about refrigerator and room temperature logs for reagents; in microbiology it may involve incubator CO2 and temperature; in molecular it may be freezer storage for extracted nucleic acid. A plan of correction that copies generic language across departments misses the actual condition the surveyor saw.
How to use these examples
- Find the tag on your own 2567 and read its guidance in Appendix C.
- Compare the finding pattern here with the evidence in your citation, noting scope (one analyte, one shift, or every test system).
- Build a correction with the four elements CMS requires: affected patients, look-back for others, systemic change, monitoring.
- Check real published 2567s in QCOR for how the tag is written up in your state.
Want a second set of eyes on your 2567?
Sound Lab Consultants helps laboratories in every specialty read their findings, find root causes and build corrections that hold up at revisit.
Related guides on soundlabconsult.com:
Frequently asked questions
What are CLIA 2567s?
Form CMS-2567 statements of deficiencies issued after CLIA laboratory surveys. Each lists the D-tags cited, the regulation and the evidence, with the laboratory's plan of correction.
Are the examples on this site real?
No. They are illustrative composites showing common finding structures. Real published 2567s can be searched in CMS's QCOR system.
Which CLIA deficiency is cited most often?
In CMS's 2024 data, D5413 (criteria for storage, test system operation and reporting conditions, 42 CFR 493.1252(b)) was cited at 818 laboratories.
Do accredited labs receive CMS-2567s?
Accredited labs are routinely inspected by their accreditor on its own forms. CMS validation and complaint surveys of accredited labs use the CMS-2567.